TM80 Telemetry Monitor Service Manual 7-7
Item Alarm Message Possible cause Solution
NIBP
NIBP Excessive
Motion
Patient’s arm moves too
much.
Check the patient’s
condition and reduce
the patient motion.
NIBP Cuff
Overpressure
The NIBP airway may be
occluded.
Check the airway and
measure again.
Cuff or Airway
Leak
The NIBP airway may
leak air.
Verify that the cuff is
properly connected.
2. Verify that the
airway does not leak
air.
NIBP Timeout Time is out.
The measurement time
is over 120 seconds.
Check the patient’s
condition and NIBP
connections.
Replace the cuff.
NIBP Cuff and
Patient Mismatch
The cuff type applied
mismatches the patient
category.
Check the patient’s
category.
Replace the cuff.
Intervals Not Set The interval in Sequence
mode is not set.
Set the intervals.
NIBP-S
Overrange
The measured NIBP
value is not within the
Contact your service
personnel.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals