EFA-200/-300 6 Operating modes
6-11
Note: The displayed field exposure takes into account all signals in the
fixed frequency range setting and the shape of the selected limit
curve. Depending on the selected limit curve, the basic display can
range up to 4%.
No. Function
1 Field quantity, in this case B-field
2 Measurement axis shown as suffix;
no suffix in this example = ISOTROPIC measurement
3 Digital display of measured value
The measured value is displayed as a % of the field
strength limit defined in the limit curve.
4 Result validity indicator:
• No icon: Computation of measured value complete
• “■”: Note
Internal calculation of result in progress, duration 1 sec.
• “ “: Warning
Measurement range exceeded
The selected range is unsuitable for measuring the field
under observation.
5 Display field F4: Display of fixed frequency range setting.
6 Display field F3: Display of selected digital display mode
7 Display field F1: Name of the limit value curve selected in
the
STANDARDS menu.
8 Bargraph: Analog display of measured value
Display range: 0.2 % to 200 %
in logarithmic steps of 1, 2, 5
Table 6-4 Exposure STD mode screen evaluation