1. INSTALLATION
REF : M22/UM/EN/015
Page 12/105
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
1.5 CONNECTION, CHANGE AND PRIMING REAGENTS
MYTHIC 22 works exclusively with the reagents described in section 4.3. Orphee’s or its
local representative will not be responsible for the quality of the results and for the
maintenance of the instrument if other commercial reagents are used.
Warranty does not cover any defect, malfunction or damage due to failure to use the
inappropriate reagents or chemicals specified for the product.
MYTHIC 22 can be configured in specific Reagent mode with encrypting in order to be used
only with Orphee’s reagents. Please contact your Orphee’s representative.
1.5.1 Connection
Lyse reagent and cleaning solution:
Before handling the reagents, read carefully their
specifications described in section 4.3.
- Remove the door on the left side of the instrument.
- Put the reagent bottles in the dedicated location.
- Remove the caps of the bottles.
- Tighten the red caps on the Lyse bottle (red sticker) and the
blue one on the cleaning solution bottle (blue sticker).
Diluent and waste:
- Connect the diluent tube (male connector) on the outlet on the bottom
and tighten the cap on the diluent container.
- To use a 20-liter diluent container adds the tubing straw adaptor
supplied with the installation kit.
- Connect the waste tube (female connector) on the outlet on the top
and tighten the cap on an empty container.
Do not modify the type and the length of the diluent
and waste tubes.
The diluent must be placed at the same level
as the
MYTHIC 22.
Collect the waste in a container preferably placed on
the floor and treat it in compliance with your local
legislation.
Example of neutralization procedure usually used in a laboratory for waste material:
For 20 liters of waste, produced by the MYTHIC 22, add 50 ml of a solution of sodium hydroxide (NaOH)
at 200g/l, mix the container, add 100ml of bleach at 36°C, mix the container, wait one hour before empty
it.
Only use this procedure if there is no specific local legislation for the treatment of the
waste.