3. INSTRUMENT SET UP
REF : M22/UM/EN/014
Page 38/105
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
4.2 PHYSICAL SPECIFICATIONS
General:
Ambient temperature: from 18 to 34°C (25°C max for reagents).
Relative Humidity: 80% maximum at 32°C.
Storage temperature: -10 to 50°C.
If the MYTHIC 22 has been stored at a temperature lower
than 10°C, it must stay at room
temperature for 24 hours before switching it on.
If the room temperature change by more than 10°C during the working day, the MYTHIC 22
must be re calibrate.
INSTRUMENT:
Dimensions: Height: 350 mm (approx.)
Width: 250 mm (approx.)
Depth: 340 mm (approx.)
Weight: 12kg (approx.)
Power supply Input: 24V 5A DC
Electric consumption: During cycle: 75 VA (-30% +10%)
Stand By: 20 VA (-30% +10%)
Maximum: 110 VA (-30% +10%)
Display: Color LCD 240*320
Mode portrait
Retro-lighted
Barcode (option): Barcode reader: C39/ Barcode / 2 interleaved 5
Memory capacity: > 500 Files (Demographics, results and histograms) in the internal memory
> 60 000 files in a 512 Kbit USB key
QC: 6 levels (100 Files per level)
Connection: RS 232C serial cable
Ethernet (TCP/IP)
Protocols: CP_M18_22
Station validation Mythic (SVM software)