Precision (Reproductibility)*
The Mythic 22 was initially calibrated with Myt-CAL calibrator (lot PLUS065, expiry date 14-05-2015).
Three levels of Myt-5D control material (lot MYT 506, expiry date 05-08-2015) were run in duplicate twice
daily for 20 days. The results were used to quantify within run precision, SD of the run means, SD of the
daily means and the total imprecision with the NCCLS EP-05 A2 guidelines.
MYT506 Normal
0.16
0.05
0.08
0.18
MYT506 Normal
0.03
0.02
0.02
0.04
MYT506 Normal
0.07
0.09
0.07
0.13
MYT506 Normal
0,21
0.73
1.90
2.05
MYT506 Normal
7,04
8.29
5.26
12.08
CV%
means
means
imprecision
MYT506 Normal
2,0
0,7
1,0
2,4
MYT506 Normal
0,7
0,4
0,5
0,9
MYT506 Normal
0,5
0,7
0,5
1,0
MYT506 Normal
0,6
2,1
5,4
5,8
MYT506 Normal
2,8
3,2
2,1
4,7
* Source 510K submission # K030799