MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 1/105
mythic 22
User’s Manual

Table of Contents

Questions and Answers

  • K
    Kelly GreenAug 18, 2025
    What to do if Orphee mythic 22 Laboratory Equipment shows 'No result' and 'No HGB & WBC'?
    • M
      Melody LeeAug 18, 2025
      If your Orphee Laboratory Equipment displays 'No result' and you suspect it's due to 'No HGB & WBC', begin by verifying the lyse level. Next, ensure the lyse tubing is correctly connected to the WBC bath. Lastly, inspect valve number 9.
  • X
    xjohnsonAug 20, 2025
    What to do if Orphee Laboratory Equipment shows 'No result' and 'No HGB'?
    • A
      Amanda VasquezAug 21, 2025
      If your Orphee Laboratory Equipment shows 'No result' and the cause is 'No HGB', verify the lyse level, ensure the correct lyse tubing connection to the WBC bath, and check valve number 9.
  • S
    Samantha RobertsAug 23, 2025
    How to fix 'Rejection Flag INS_H' on Orphee mythic 22?
    • N
      natalie05Aug 23, 2025
      To address a 'Rejection Flag INS_H' on your Orphee Laboratory Equipment, close the door and perform a new Start Up cycle. Repeat the Start Up cycle again.
  • S
    skinnerjustinAug 26, 2025
    What to do if Orphee Laboratory Equipment shows 'No result' and 'No HCT & PLT'?
    • J
      Joshua DuncanAug 26, 2025
      If your Orphee Laboratory Equipment displays 'No result' and the cause is 'No HCT & PLT', check the bath wires. Then, perform a cleaning cycle, and if the issue persists, follow with a bleach cycle.
  • O
    orozcomeganAug 27, 2025
    What to do if Orphee mythic 22 shows 'No result' but 'HGB OK'?
    • R
      Richard WilliamsAug 27, 2025
      If your Orphee Laboratory Equipment displays 'No result' but 'HGB OK', check the bath wires. If the issue persists, perform a cleaning cycle followed by a bleach cycle. If still unsuccessful, change the aperture block.
  • J
    jasmine35Aug 30, 2025
    What to do if Orphee mythic 22 Laboratory Equipment shows 'No result or bad stability' but 'HGB & WBC OK'?
    • J
      James BellAug 30, 2025
      If your Orphee Laboratory Equipment displays 'No result or bad stability' but 'HGB & WBC OK', perform a back flush cycle, and if the issue persists, follow with a bleach cycle. Check valves number 12, 13, 16 & 17. Also, verify the reagent and enclosure temperature and check the preamp connection wires.
  • A
    Ashley MorganSep 1, 2025
    What to do if Orphee Laboratory Equipment shows Rejection Flag INS_H?
    • R
      Robert ThompsonSep 1, 2025
      If your Orphee Laboratory Equipment shows a rejection flag INS_H, close the door and perform a new Start Up cycle. Repeat the Start Up cycle if needed.
  • K
    kristinabensonSep 3, 2025
    How to fix bad stability in Orphee mythic 22?
    • A
      Alice WrightSep 4, 2025
      If your Orphee Laboratory Equipment has bad stability, perform a Back flush and a Cleaning Cycle, followed by a Bleach cycle if the issue persists. Also, check the level bubble flow in the WBC bath during the run cycle, and verify the reagent and enclosure temperature.
  • B
    Beth JohnsonSep 6, 2025
    What to do if Orphee Laboratory Equipment shows no result but HGB is OK?
    • J
      Joseph StevensSep 6, 2025
      If your Orphee Laboratory Equipment shows no result, and HGB is OK, check the bath wires. If the issue persists, perform a Cleaning Cycle and then a Bleach cycle. If still unsuccessful, change the aperture block.
  • M
    Megan PerezSep 8, 2025
    What to do if Orphee mythic 22 Laboratory Equipment shows no result, no HCT & PLT?
    • X
      xspearsSep 8, 2025
      If your Orphee Laboratory Equipment shows no result and there is no HCT & PLT, check the bath wires. Perform a Cleaning Cycle and then a Bleach cycle if unsuccessful.

Summary

1. INSTALLATION

Guidance and manufacturer’s declaration - electromagnetic emissions.

Manufacturer's declaration on electromagnetic emissions for the MYTHIC 22.

Guidance and manufacturer’s declaration - electromagnetic immunity.

Manufacturer's declaration on electromagnetic immunity for the MYTHIC 22.

1.1 UNPACKING

Procedures for safely unpacking the MYTHIC 22 instrument.

1.2 INSTALLATION CONSTRAINTS

Requirements for the installation location and environment of the MYTHIC 22.

1.3 ELECTRICAL CONNECTIONS

Details on connecting the MYTHIC 22 to power and other devices.

1.4 PRINTER CONNECTION

Instructions for connecting the printer to the MYTHIC 22.

1.5 CONNECTION, CHANGE AND PRIMING REAGENTS

Steps for connecting, replacing, and priming reagents for the MYTHIC 22.

1.6 TRANSPORTATION AND STORAGE

Guidelines for transporting and storing the MYTHIC 22 instrument.

DANGER

WARNING! RISK OF DANGER!

Explains hazard warning symbols and their meanings.

NOTA

Explains important additional information symbol.

2. GENERAL OVERVIEW

2.1 GENERALITIES

General description of the MYTHIC 22 hematology analyzer.

2.2 OVERVIEW

Overview of the main parts of the MYTHIC 22 instrument.

2.3 MAIN PART DESCRIPTION

Detailed description of the main components of the MYTHIC 22.

3. INSTRUMENT SET UP

3.1 USER’S IDENTIFICATION

Procedure for logging in and identifying users on the MYTHIC 22.

3.3 SET UP

Configuration options for language, printer, and system settings.

3.4 ADVANCED SET-UP

Advanced configuration menus for biologists and technicians.

4. SPECIFICATIONS

4.1 ANALYTICAL SPECIFICATIONS

Performance specifications for analysis, including throughput, sample volume, and principles.

4.3 REAGENTS SPECIFICATIONS

Specifications, application, and precautions for reagents.

4.4 ANALYTICAL LIMITATIONS

Examples of specimens causing errors and actions to take.

5. SAMPLE ANALYSIS

5.2 START UP

Procedure for starting up the MYTHIC 22 instrument.

5.5 PREPARATIONS BEFORE ANALYSIS

Procedures for sample collection and preparation before analysis.

5.6 ANALYSIS

Detailed procedure for running sample analysis.

5.7 RESULTS

Display and interpretation of analysis results and flags.

5.11 STAND BY AND SHUT DOWN

Procedures for putting the instrument in standby or shutting it down.

6. QUALITY CONTROL

6.2 QC

Accessing the quality control menu and managing lots.

6.2.2 Run control blood

Procedure for running control blood samples.

6.3 REPEATABILITY

Performing and analyzing repeatability tests.

7. CALIBRATION

7.1 CALIBRATION

When and how to perform instrument calibration.

7.2 TARGET VALUE MODIFICATIONS

Modifying target values and creating new batches for calibration.

8. TECHNOLOGY

8.1 DETECTION PRINCIPLE

Explanation of the impedance technique for cell counting.

8.1.2 Five-part diff measurement

Explanation of the optical detection system for five-part differential analysis.

8.5 FLAGS

Explanation of various flags indicating potential issues.

8.6 HYDRAULIC DESCRIPTION

Description of the instrument's hydraulic modules.

9. SERVICE

9.1 MAINTENANCE

Importance of maintenance and a detailed maintenance schedule table.

9.4 REPAIRING

Procedures for repairing the instrument.

9.5 TROUBLESHOOTING

Guide to resolving common instrument problems.

9.7 LOGS ERRORS

Viewing and printing error logs for troubleshooting.