6. QUALITY CONTROL
6.1 INTRODUCTION
Quality control allows checking the stability of the MYTHIC 22 analytical performances when operating.
The control blood must be used before its expiry date and stored according to the
manufacturer instructions for use. It must be manually well-mixed before use.
In case of no local regulation, it is recommended to run a control blood at the beginning of each working day
before running sample.
In case of exceeding the tolerances indicated on the blood control result sheet, it is recommended to
perform a cleaning cycle (see section 9.1.2) followed by a STARTUP cycle.
If result still exceeding the tolerances indicated on the blood control result sheet, it is recommended to
perform a calibration (see section 7).
6.2 QC
MYTHIC 22 stores in memory up to 100 results per lot, for 6 different lots. Results of each lot can be
viewed in tables and Levey-Jennings graph.
- Press to have access to the quality control menu.
- Quality control display presents the identification of the lots.
- The last active lot is labeled with a dark dot on the left lot.
- To choose another lot, press on the wished lot.
- The key allows the
modification of the identification and the target values.
- The key allows restoring the data from a USB key.
- The key allows:
. To view the result table.
. To perform quality control analysis.
- To print, send or save the targets, results and graphics press .