9.4.4 Baths o-ring replacement
Wear rubber gloves and wash hands with a disinfectant after completion of work.
NOTA: This operation can be performed without dismantling the complete bath block. Tubes and connectors
can stay.
-
Before performing the operation, drain the baths and
dismantle the counting bath block (see section 9.4.3).
- Replace the o-ring and/or the complete bath block.
- The counting baths o-
rings need to be greased with silicon
grease present in the maintenance box.
REASSEMBLY PROCEDURE:
To reassemble the unit, carry out the various operations in the
opposite direction.
Place the seal on the counting bath gently not to wound it.
9.4.5 Aperture counting head replacement
Wear rubber gloves and wash hands with a disinfectant after completion of work.
- Before performing the operation, drain the baths and
dismantle the counting bath block (see section 9.4.3).
-Rotate the counting head of ¼ tour.
-
With the help of the provided tool in the maintenance kit,
unscrew the screw and remove the connection.
NOTA: This operation can be performed without dismantling
the complete bath block.