3.4.3 Analysis options:
- OPTIONS box:
o ALWAYS ID and ALWAYS PID: To run a sample, user mandatory
needs to enter a ID and/or a PID.
o RUO: With tag the PCT, PDW and PLCR parameters are displayed,
printed and send.
o US MODE: The Research Use Only message is printing below the
printing report.
o ABSOLUTE DIF: With tag absolute values for sub-populations of
leucocytes are displayed. In the other case, percentages are displayed.
o STARTUP ALARMS: With tag, the message SU_F “Startup Failed”
appears below the printing report.
o QC ALARMS: The message QC_F “QC failed” appears below the
printing report when the QC result is out of tolerance or expired (see
section 8.5.5); The message QC_ND “QC not done” appears below the
printing report when it is not run (see section 8.5.5).
- : Gives a choice of three unit systems: Standard,
International System, and mmol.
- Barcode Reader box: The reading of the bar code is dedicated to the PID
or ID or SID.
- BOLD DISPLAY box: display and print in bold-faced type the different
choices in this box.
Once modifications are done, press either to validate your choices or to exit keeping the
previous setting.
3.4.4 Lab. parameters:
- Select the blood type in the combo box then
press:
to adjust the normal and panic limits (see section
3.4.4.1).
to adjust the parameters thresholds (see section
3.4.4.2).
to adjust the flags level (see section 3.4.4.3).
to adjust the corrections factors (see section
3.4.4.4).NOTA: No correction factors with type STANDARD.
- To enter a new blood type, press
NOTA: The name of the first type STANDARD cannot be change.
- Press to print the blood type set up.
- allow to print all the blood type set up (about 20 pages are
printed).
- only the blood type in the combo box is printed.