Copyright© Orphée SA. All Rights Reserved.
1.5.6 Waste setting:
- Enter only the capacity of the container.
- After replacement of the waste container press to
reset to initialize the waste calculation.
1.6 TRANSPORTATION AND STORAGE
Storage temperature: -10°C to +50°C.
If the MYTHIC 22 has been stored at a temperature less than 10°C it must be left at room temperature for
24 hours.
Before transportation outside the laboratory, perform a complete cleaning with a disinfectant in compliance
with the local legislation. Please use the CLEANOUT procedure, for details contact your local ORPHEE’S
representative.
Transportation requirements:
• Keep the parcel, packaging and box containing the device DRY at all times.
• The Mythic 22 is very fragile, has breakable parts, and must be handled cautiously.
• The box containing the device must be handled put down or moved keeping the designated face up at
all times. It must not be tilted.
• Don’t pile up more than TWO boxes containing devices, either for transportation or long term
storage.
Perform a calibration, section 7 and a quality control, section 6, then the
MYTHIC 22 WILL BE READY TO OPERATE.