Copyright© Orphée SA. All Rights Reserved.
- Wait for the information window and then,
- Open the door on the right hand
section 1.1.3).
- Put 4 ml of Mythic 18-22 FLUSH - CLEANER
(Ref. HM18-011-025) in each counting chamber
the side door.
- Press the button in the window l
center.
- MYTHIC 22
perform a maintenance cycle of the
apertures followed with a standby mode during 10 min.
- After 10 min. MYTHIC 22 rinses all of the elements.
Perform a blank cycle before to run
analytic cycle.
Wear rubber gloves and wash hands with a
disinfectant after completion of work.
9.1.3 Piston greasing
This simple operation keeps the state of the pistons and the pistons o-rings in perfect state which
allow extending the lifetime between the pistons o-rings exchange.
- To have access to the hydraulic cycle, press from the
MAIN MENU, then press .
-
Press to put the piston of the syringe module in the
greasing position.
To avoid any risk of injuries, w
ait complete end of piston
movement before to perform the greasing.
NOTA
: The greasing frequency must be increased proportionally to the daily
workload (see maintenance table 9.1.1).