9. SERVICE
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 84/105
9. SERVICE
The quality of the results and the reliability of the MYTHIC 22 are directly linked to the strict respect of
the maintenance hereafter described.
To perform the maintenance and the repair described in this section, it is mandatory to have
received adequate training, use the Orphee spare parts accordingly, to wear rubber gloves
and wash hands with a disinfectant after completion of work.
9.1 MAINTENANCE
9.1.1 Maintenance table
NOTA: This table is made for an average number of 50 samples per day. For more, please increase
proportionally the number of maintenances.
MAINTENANCE
DAILY
WEEKLY
QUATERNALY
ANNUALLY
2 YEARS OR
25’000 SAMPLES
User
Tech
User
Tech
User
Tech
User
Tech
Tech
Reagents level
X
Start Up
X
Automatic cleaning
X
Concentrate cleaning
X
Shut down
X
Cover cleaning
X
Pistons greasing
X
X
Motor screw greasing
X
Needle o-ring replacement
X
Counting baths O-ring
replacement
X
Counting heads O-ring
replacement
X
Syringes O-rings
replacement (frequency is
directly related to pistons
greasing)
X
- Reagents level control: see section 5.3.
- Start up: see section 5.2.
- Automatic Cleaning: The cleaning is automatically performed by the MYTHIC 22 following the set up
defined by the Tech (see section 3.3.6) (the standard value is cleaning made every25 samples).
Change the frequency of the cleaning of the MYTHIC 22 (mean decrease the number, 15 for ex.) in case of
analyzes of pathological samples or if there are too many flags and too often.
- Concentrated cleaning: see the following section 9.1.2.
- Shut down: see section 5.11.
- Cover cleaning: Clean the cover above and around the sampling needle with a wet paper with a disinfectant
to eliminate the blood trace.
- Piston greasing: see section 9.1.3 please note that the correct and frequent greasing of the syringes
pistons extend the time between the syringes O-rings replacement.
NOTA: Maintenance information’s are registered in a log (see section 5.9).