4. SPECIFICATIONS
4.1 ANALYTICAL SPECIFICATIONS
Throughput: > 45 samples/hour
Sample Volume: 15,7 µl
Measurement Principle: WBC/RBC/PLT: Impedance
Five part diff: Optical flow cytometry
Hemoglobin: Spectrophotometry at 555 nm
Hematocrit: Volume integration
Linearity:
Linearity is measured with linearity sample performed four times for each level.
Reportable range:
Within the reportable range, the results are flagged with a D to indicate that it is necessary to re-dilute
and re-run the sample.
Repeatability:
Calculated with 20 runs of a fresh whole blood sample performed on a commercial MYTHIC 22.
×
Calculated method: CV = SD
X
Σx²- (Σx)²
SD = N
N - 1
Please note that this table gives the acceptable references of CV's used in calibration process