3. INSTRUMENT SET UP
REF : M22/UM/EN/014
Page 36/105
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
Carry-Over:
For each parameter, we perform 3 runs from a high concentration sample followed by 3 runs without sample.
WBC
RBC
HGB
PLT
High concentration value
145
×
10
3
/mm
3
8.40
×
10
6
/mm
3
20 g/dL
2300
×
10
3
/mm
3
Measured carry-over (%)
0.68
0
0
0.04
Maximum carry-over (%)
< 1,0
< 1,0
< 1,0
< 1,0
The percentages of carry-over inter samples is calculated with the following formula:
(Low value cycle 1) - (Low value cycle 3)
Carry-over =  × 100
(High value cycle 3) - (Low value cycle 3)
Accuracy:
Correlation was done with one commercial instrument (HORIBA ABX PENTRA 120 ®) and with about 100
normal blood samples (without alarm).
PARAMETERS
N
R (%)
WBC (10
3
/mm
3
)
127
0,997
LYM (%)
113
0,989
MON (%)
113
0,935
NEU (%)
113
0,988
EOS (%)
113
0,950
BAS (%)
113
0,187
RBC (10
6
/mm
3
)
127
0,991
HGB (g/dL)
127
0,997
HCT (%)
127
0,984
MCV (fL)
127
0,947
RDW (%)
127
0,746
MCH (pg)
127
0,963
MCHC (g/dL)
127
0,151
PLT (10
3
/mm
3
)
127
0,990
MPV (fL)
127
0,890
n.Σxy – Σx. Σy
Calculated method: r = 
(n. Σx²-(Σx²))(n. Σy²- (Σy²))
PENTRA 120 ® is a registered trademark of HORIBA ABX.