1.5.2 Priming
When first starting the MYTHIC 22, it is necessary to perform a complete prime of the fluidic circuit.
Before starting, be sure that all the reagent and waste tubes are properly connected.
The reagents must be stored minimum 24 hours at room temperature before use.
Priming procedure :
Switch on:
- Connect the power supply block.
- Press the ON/OFF button.
- The cycle LED turns red. No cycle can be performed before it turns
green.
- The information window could stay up to 3 mn to enable the update of all
the files.
Login:
- The operator’s identification display appears.
- Enter the user’s identification, the password (see section 3.1) and press
to validate.
- AUTONOMY (run) indicates the number of samples (runs) you can
perform (calculated with the smaller quantity of reagents).
- If this window appears, it means that several
results in memory have not been sent before
the MYTHIC 22 was switched off.
- Press YES to send them immediately or press
LATER to wait at another time or NO if you do
not want to send them.
- No USB key is available, connect an USB key then press YES or see section 3.4.7
to change the archive mode.