9.5.2 Other problems
Diluent leaks around the
needle during the run cycle.
Check the rinsing needle block (presence of clots) and clean
it if necessary see section 9.4.2.
Check the power supply connection wires
Check the switch on button located on the power supply
block.
Check the level diluent and if the tubing is pinched.
Check if the diluent is at the same level than the analyzer.
ID and/or PID typing
impossible.
ID and/or PID are mandatory (see section 3.4.3).
OUT OF RANGE
INS_T.
Check the origin of the problem of temperature in section 9.3.
Trigger temperature values:
Ambient: 17°C <--> 37°C,
Enclosure: 33°C <--> 40°C,
Reagents: 34°C <--> 45°C.
DEFAULT or Flag INS_P.
- drain chamber default:
check the tubing connection of the
fluidics.
- counting vacuum default: perform a piston greasing (see section
9.1.3) and check the tubing connection of the fluidics,
For all occurrences check also the level of the reagents.
Check the paper.
Check the electrical connection.