9. SERVICE
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 97/105
9.5.2 Other problems
ORIGIN
PROBLEMS
SOLUTIONS
MYTHIC
Diluent leaks around the
needle during the run cycle.
Check the rinsing needle block (presence of clots) and clean
it if necessary see section 9.4.2.
No starting.
Check the power supply connection wires
Check the switch on button located on the power supply
block.
All results bad.
Check the level diluent and if the tubing is pinched.
Check if the diluent is at the same level than the analyzer.
No display.
Check the flat cable.
ID and/or PID typing
impossible.
ID and/or PID are mandatory (see section 3.4.3).
Message : TEMPERATURE
OUT OF RANGE
or flag
INS_T.
Check the origin of the problem of temperature in section 9.3.
Trigger temperature values:
Ambient: 17°C <--> 37°C,
Enclosure: 33°C <--> 40°C,
Reagents: 34°C <--> 45°C.
Message: CYCLE: PRESSURE
DEFAULT or Flag INS_P.
Cases of occurrence:
- drain chamber default:
check the tubing connection of the
fluidics.
- counting vacuum default: perform a piston greasing (see section
9.1.3) and check the tubing connection of the fluidics,
For all occurrences check also the level of the reagents.
Printer
No printing.
Check the paper.
Check the electrical connection.
Bad printing.
Check the black ribbon.