5.11 STAND BY AND SHUT DOWN
NOTA: It is recommended to switch off the MYTHIC 22 if it will be not used for next several hours.
- From the main menu, press . MYTHIC 22 automatically
performs a shut down cycle.
- All the hydraulic circuits are rinsed and then cleaned with the cleaning
solution.
- At the end of the cycle MYTHIC 22 automatically stop.
- Shut Down can be automatically performed after a setting time (see
section 3.4.6).
NOTA: After a shutdown, it is impossible to perform an analytical cycle without launching a startup
cycle. (See section 5.2)
Station Validation Mythic (SVM)
SVM is windows application that allows connecting up to 4 Mythics via serial cable.
Worklist with complete demography, validation, modification of results, archiving patients,
calibration, QC, repeatability, made statistics on XB, Levey Jennings, delta check, daily mean and
comments. Please contact your Orphee’s representative for further information.
MYTHIC 22 must stay at least,
with cleaning
solution (after a shutdown), for
three hours
every 24 hours for cleaning purpose.