8. TECHNOLOGY
REF : M22/UM/EN/014
Page 80/105
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
WBC CURVE FLAGS:
This curve is done by the WBC channel measurement and does not appear on the normal results screen. The
flags L1 and L5 complete the flags N2 and IC.
L1 = Presence of platelets
aggregate, erythroblast or small
lymphocytes at the left of CL1.
L5 = Presence of immature cells at the right of CL5.
8.5.4 Erythrocyte and HGB Flags
R1 = Abnormal quantity of microcytes at the left side of CR1.
R2 = Abnormal quantity of macrocytes at the right side of CR2.
HGB: * means bad measure of the HGB blank.
8.5.5 Platelets Flags
P1 = 0 to CP1
: Presence of abnormal quantity of debris
or small cells.
P2 = P to CP2: Presence of schizocytes.
P3 = CP3 to CP3-2: Presence of microcytes.
8.5.6 QC Flags
QC_F: QC Failed means that the results are outside the limits entered for one or several parameters (see
section 6.2.1).
QC_ND: QC Not Done means that the Quality Control has not been done or that the analysis are done with
QC results outside the limits (see section 6.2.1).
8.5.7 STARTUP Flags
SU_F: STARTUP Failed means that the results are outside the limits for one or several parameters (see
section 5.4).
SU_ND: STARTUP not done means that the Start Up has not been done or that the analysis is done with
results of the Start Up outside the limits (see section 5.4).