This equipment needs special precautions regarding general requirements for safety.
Guidance and manufacturer’s declaration – electromagnetic emissions.
The MYTHIC 22 is intended for use in the electromagnetic environment specified below. The customer or
the user of the MYTHIC 22 should assure that it is used in such an environment.
Electromagnetic environment - guidance
The MYTHIC 22 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Voltage fluctuations/flicker
emissions
Guidance and manufacturer’s declaration – electromagnetic immunity.
The MYTHIC 22 is intended for use in the electromagnetic environment specified below. The customer or
the user of the MYTHIC 22 should assure that it is used in such an environment.
Electromagnetic environment - guidance
discharge (ESD)
IEC 61000-4-2
±8 kV air
Complies
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30 %.
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
Complies
Mains power quality should be that of a typical
commercial or hospital environment.
IEC 61000-4-5
±2 kV common mode
Complies
Mains power quality should be that of a typical
commercial or hospital environment.
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
(>95 % dip in
U
T) for 0,5 cycle
40 %
U
T
(60 % dip in
U
T) for 5 cycles
70 %
U
T
(30 % dip in
U
T) for 25 cycles
<5 %
U
T
(>95 % dip in
U
T) for 5 sec
Complies
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
MYTHIC 22 requires continued operation during power
mains interruptions, it is recommended that the MYTHIC
22 be powered from an uninterruptible power
supply or a battery.
(50/60 Hz)
magnetic field
IEC 61000-4-8
Complies
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.