6. QUALITY CONTROL
MYTHIC 22
Copyright© Orphée SA. All Rights Reserved.
Page 67/105
- The key allows having access to the repeatability
display. (see section 4.1) repeatability table.
This screen allows carrying out a test of repeatability on all the parameters
measured by the MYTHIC 22.
- Dry the needle with a clean paper towel before each aspiration.
- Present the sample under the sampling needle and press in the trigger
located behind the needle.
- The cycle LED located at the top of the needle flickers alternatively
of red with green. When it becomes red the tube can be removed.
- Repeat the operation as many times as desired (maximum 20 runs).
- The results are progressively sent on line in the table.
- Statistical calculations are automatically carried out with each run.
- The cursor located between the two tables enables to send the other
results (the results of statistical calculations move at the same time).
- The column allows to validate or to unselect the result.
Please refer to the chapter 4.1 Analytical specifications, repeatability table, for checking values.
- The key allows carrying out the following actions:
o Print the table.
o Send the results.
o Delete the results.