EasyManua.ls Logo

Oval ALTI mass CA00A - Page 130

Oval ALTI mass CA00A
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
L
--
740
--
28
--
E
10
■ KCs explosionproof
The cable gland is not attached to the electric wire connection port. Please prepare it at customers side.
In addition, the specifications for usable cable gland are as follows;
〈Caution〉
Use the cable gland with minimum of 5 screw threads engaged when connecting the cable gland to
the converter electric connection part.
Seal off properly not to make the infiltration of rain water and dust from cable, rubber packing, screw
  tightening part and so on.
Usable Cable Gland
Conformance Certification : The cable gland should be the one which certified for the explosionproof in
South Korea.
Explosion-proof Notation : Ex d IIC Gb
Screw Joint Part : G3/4
IP Protection Class : IP66/67 and higher class
Operating Temperature Range : 40 to +60
Integral Type Converter