CIRAS-3 Operation Manual V. 1.09 168 support@ppsystems.com
It is taking longer than normal for my CO2r levels to settle
Under settings your CO2 Reference is likely set to “Fixed analysis air”. Change to “Fixed
reference air”.
I am unable to achieve a Max C calibration above 2000 ppm and it is failing
Your CO
2
cartridge is low and needs to be replaced.
Your desiccants are exhausted and require replacement. Replace all desiccants.
H
2
O Measurement and Control
My H2Or levels are reading higher than expected
There may be a leak somewhere in the gas circuit allowing external ambient air into the system.
Usually this is associated with the absorber columns and the result of a recent desiccant change.
Check that both absorber column assemblies are firmly fitted to the CIRAS-3 console and check
all “O” rings to ensure that they are all firmly seated and not pinched off or cracked.
Exhausted H
2
O scrubber in the Zero column. Check the Drierite and replace.
My H2Or levels are reading lower than expected
Exhausted H
2
O scrubber in the Zero column. Replace the Drierite and Molecular Sieve.
My H2Or or H2Oa values are fluctuating
First I would check your desiccants and change out if exhausted.
There may be a leak somewhere in the gas circuit allowing external ambient air into the system.
Usually this is associated with the absorber columns and the result of a recent desiccant change.
Check that both absorber column assemblies are firmly fitted to the CIRAS-3 console and check
all “O” rings to ensure that they are all firmly seated and not pinched off or cracked.
Your reference or sampling pump might be dirty or dusty and require cleaning or replacement.
See Servicing the Air Sampling Pump (Reference & Analysis). On page 181
. Another quick trick
is to simply swap the pumps to see if the problem follows the pump.
Dirty air supply unit hydrophobic filter. See Air Filters on page 183 for more details.
My H
2
O readings are drifting
Exhausted H
2
O scrubber in the Zero column. Replace the Drierite and Molecular Sieve.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PP Systems CIRAS-3 and is the answer not in the manual?

PP Systems CIRAS-3 Specifications

General IconGeneral
BrandPP Systems
ModelCIRAS-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals