CIRAS-3 Operation Manual V. 1.09  56  support@ppsystems.com 
 
PAR Measurement 
The PLC3 leaf cuvettes also features two ways of measuring light in the 400-700 nm wavelengths.  All 
PLC3s include a cosine-corrected external PAR sensor (PARe) for measurement of ambient PAR 
(Photosynthetically Active Radiation).  This reading will be most reliable with the cuvette held on a 
horizontal plane relative to the ground.  
PLC3 Universal - includes two mini-silicon photodiode sensors beneath the cuvette window (PARi).   The 
silicon photodiode sensors are close to the leaf plane and used to average the irradiance beneath the 
cuvette window.  Irradiance is somewhat attenuated (approx. 10% attenuation) by the window.  This 
affects the amount of light reaching the leaf if the light source is ambient sunlight, but not if our LED light 
unit is the source.  This is because the internal PAR sensors are on an electronic feedback loop with the 
light unit, so the desired light intensity entered by the user is always achieved. 
PLC3 Conifer and Narrow – includes a single, cosine corrected PAR sensor beneath the cuvette window 
(PARi).  It is used to give an indication of the PAR inside the leaf cuvette and as an electronic feedback 
loop to control light intensity when our light unit is used. 
 
PLC3 Universal Leaf Cuvette 
 
PLC3 Conifer and Narrow Leaf Cuvette 
 
 
TIP 
The PARi sensor will normally read about 10% lower than the PARe sensor due to attenuation of the 
cuvette window.