CIRAS-3 Operation Manual V. 1.09  170  support@ppsystems.com 
 
I am unable to achieve my Tcuv set point 
•  You are working outside the temperature control limits of 0-45 
o
C. 
•  The set point is not achievable because it is outside the temperature control range of ~10 
o
C 
below ambient to +15 
o
C above ambient. 
•  More than likely you are unable to cool the chamber to your set point as a result of high incident 
light or your batteries are getting low. 
I am unable to achieve my Tleaf set point 
•  You are working outside the temperature control limits of 0-45 
o
C. 
•  The set point is not achievable because it is outside the temperature control range of ~10 
o
C 
below ambient to +15 
o
C above ambient. 
•  More than likely you are unable to cool the chamber to your set point as a result of high incident 
light or your batteries are getting low. 
 
 
 
PAR Measurement and Control 
Why does my PARi reads lower than PARe 
•  This is normal.  The PARi sensor is located inside the cuvette head and below the window and 
should therefore read approximately 10% lower than PARe. 
•  Make sure that nothing is obstructing the PARi sensors. 
I am using the light unit but I am unable to control light intensity 
•  Check to make sure that the electrical connection is in place. 
•  Go to Settings and make sure that you have the proper light source selected 
My PARe readings do not seem accurate 
•  The sensor likely requires recalibration.  Check the calibration against another PAR sensor that 
you trust and believe to be calibrated.  Recalibrate if necessary.  See PAR Calibrate on page 106
.