CIRAS-3 Operation Manual V. 1.09 179 support@ppsystems.com
Battery Pack (Aved)
Part Number 41526-1
Battery Pack (Inspired Energy)
Part Number 41535-1
We recommend that the battery is charged every night when the system is in regular use and ALWAYS
charge the internal battery(s) to full capacity prior to storing the system for extended periods. We also
highly recommend that the internal battery(s) gets charged every 3-4 months when not in use to preserve
capacity. A power supply is included with the CIRAS-3 for charging the internal battery(s) or for extended
operation. To charge the internal battery(s), connect the power supply to the “EXT PWR” plug on the
back of the CIRAS-3 main console. The indicating LED above the plug shows the charge status of the
internal battery. It will illuminate a steady green when charging and when the battery is fully charged.
The battery capacity can be observed in the lower right hand corner of the display when the system is up
and running with the measurement or graphical display on the LCD.
If you need to order a replacement battery pack please make sure that you are ordering the correct type
as highlighted earlier.
Battery Removal
If a battery needs to be removed or replaced, simply open the front battery cover and allow the door to
gently drop down. Grasp the black tab on the battery and pull it out of the compartment. See
Battery
Pack Installation on page 46 for information on installation of battery packs.
Battery Storage
Li-ion batteries do not have a “memory” so there is no need to fully discharge them after use. If the
instrument is not going to be used for weeks or months then we recommend that the internal battery
packs have at least 25-50% charge capacity before storing away.
Battery Disposal
Proper and safe disposal of Lithium-ion batteries is critical as improper disposal
may cause damage to human health or the environment. Please check your local
area for Lithium-Ion battery recycling locations to learn how you can safely dispose
of the batteries. Material Safety Data Sheets (MSDS) are available from the battery
manufacturer if required. Please contact PP Systems for more information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PP Systems CIRAS-3 and is the answer not in the manual?

PP Systems CIRAS-3 Specifications

General IconGeneral
BrandPP Systems
ModelCIRAS-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals