Remote Commands
R&S
®
FSW
746User Manual 1173.9411.02 ─ 19
Parameters:
<Function> NONE
(reference ranges only:) the limit of the reference range is used;
Reference ranges always use the function "NONE".
SUM
sum of the two limit lines (calculated for linear powers) is used
MAX
maximum of the two limit lines is used
*RST: SUM (reference range: NONE)
Manual operation: See "Multi-Limit Calc <n>" on page 214
[SENSe:
]ESPectrum<sb>:RANGe<range>:RLEVel <RefLevel>
This command defines the reference level for a SEM range.
In case of high speed measurements, the reference level has to be identical for all
ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Parameters:
<RefLevel> Reference level.
Refer to the data sheet for the reference level range.
*RST: 0 dBm
Manual operation: See "Ref. Level" on page 212
[SENSe:]ESPectrum<sb>:RANGe<range>:SWEep:TIME <SweepTime>
This command defines the sweep time for a SEM range.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Parameters:
<SweepTime> Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Manual operation: See "Sweep Time" on page 212