Remote Commands
R&S
®
FSW
692User Manual 1173.9411.02 ─ 19
//Query one of the SEM results
CALCulate:MARKer:FUNCtion:POWer:RESult? CPOWer
//Switch to second Spectrum channel
INST:SEL 'Spectrum 2';*WAI
//Query channel power result
CALCulate:MARKer:FUNCtion:POWer:RESult? ACPower
11.5 Configuring and Performing Measurements
The following commands are required to configure measurements in a remote environ-
ment. The tasks for manual operation are described in chapter 5, "Measurements",
on page 110.
● Performing Measurements....................................................................................692
● Configuring Power Measurements........................................................................695
● Measuring the Channel Power and ACLR............................................................ 700
● Measuring the Carrier-to-Noise Ratio................................................................... 726
● Measuring the Occupied Bandwidth..................................................................... 727
● Measuring the Spectrum Emission Mask..............................................................729
● Measuring Spurious Emissions.............................................................................765
● Analyzing Statistics (APD, CCDF)........................................................................ 778
● Measuring the Time Domain Power......................................................................787
● Measuring the Harmonic Distortion.......................................................................795
● Measuring the Third Order Intercept Point............................................................799
● Measuring the AM Modulation Depth....................................................................801
● Remote Commands for EMI Measurements.........................................................803
● List Evaluations.....................................................................................................811
● Measuring the Pulse Power.................................................................................. 816
11.5.1 Performing Measurements
Useful commands for performing measurements described elsewhere
● INITiate:ESPectrum on page 731
● INITiate:SPURious on page 765
Remote commands exclusive for performing measurements:
ABORt.......................................................................................................................... 692
INITiate:CONMeas.........................................................................................................693
INITiate:CONTinuous..................................................................................................... 694
INITiate[:IMMediate]....................................................................................................... 694
[SENSe:]SWEep:COUNt:CURRent?................................................................................695
ABORt
This command aborts a current measurement and resets the trigger system.
Configuring and Performing Measurements