Parameters:
<Position> 0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST: 100 PCT = frequency display; 50 PCT = time dis-
play
Example:
DISP:TRAC:Y:RPOS 50PCT
Usage: SCPI confirmed
Manual operation: See "Reference Position" on page 349
See "Ref Level Position" on page 400
DISPlay[:
WINDow<n>]:TRACe:Y:SPACing <ScalingType>
This command selects the scaling of the y-axis.
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Usage: SCPI confirmed
Manual operation: See "Scaling" on page 401
11.7.4 Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here. The tasks for manual operation are described in
chapter 6.6, "Trigger and Gate Configuration", on page 418.
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.