Measurements
R&S
®
FSW
246User Manual 1173.9411.02 ─ 19
For details on transducers see chapter 9.2, "Basics on Transducer Factors",
on page 566.
Remote command: 
[SENSe:]LIST:RANGe<range>:TRANsducer on page 773
Limit Check
Activates or deactivates the limit check for all ranges.
For details on limit checks see chapter 5.7.3.2, "Limit Lines in Spurious Measure-
ments", on page 241.
"Absolute"
Signal is checked against absolute limit values
"None"
No limit check is performed.
Remote command: 
[SENSe:]LIST:RANGe:LIMit:STATe on page 771
CALCulate:LIMit<k>:FAIL? on page 987
Abs Limit Start/Stop
Sets an absolute limit value at the start or stop frequency of the range [dBm].
Remote command: 
[SENSe:]LIST:RANGe<range>:LIMit:STARt on page 771
[SENSe:]LIST:RANGe<range>:LIMit:STOP on page 771
Insert before/after Range
Inserts a new range to the left of the currently focused range (before) or to the right
(after). The range numbers of the currently focused range and all higher ranges are
increased accordingly. The maximum number of ranges is 30.
Delete Range
Deletes the currently focused range. The range numbers are updated accordingly.
5.7.4.2 Adjusting the X-Axis to the Range Definitions
The frequency axis of the measurement diagram can be adjusted automatically so that
the span of all sweep list ranges corresponds to the displayed span. Thus, the x-axis
range is set from the start frequency of the first sweep range to the stop frequency of
the last sweep range.
5.7.4.3 List Evaluation
In the "List Evaluation" dialog box, which is displayed when you select the "Evalua-
tions" button in the "Overview" or the "List Evaluation" softkey in the "Spurious Emis-
sions" menu, you configure the contents and display of the result list.
 
 
Spurious Emissions Measurement