Remote Commands
R&S
®
FSW
907User Manual 1173.9411.02 ─ 19
11.8 Analyzing Measurements (Basics)
The commands for general analysis tasks are described here.
● Zooming into the Display.......................................................................................907
● Configuring the Trace Display and Retrieving Trace Data....................................909
● Working with Markers............................................................................................928
● Configuring Display and Limit Lines......................................................................975
11.8.1 Zooming into the Display
11.8.1.1 Using the Single Zoom
DISPlay[:WINDow<n>]:ZOOM:AREA............................................................................... 907
DISPlay[:WINDow<n>]:ZOOM:STATe..............................................................................907
DISPlay[:WINDow<n>]:ZOOM:AREA <x1>,<y1>,<x2>,<y2>
This command defines the zoom area.
To define a zoom area, you first have to turn the zoom on.
1
= origin of coordinate system (x1 = 0, y1 = 0)
2 = end point of system (x2 = 100, y2= 100)
3 = zoom area (e.g. x1 = 60, y1 = 30, x2 = 80, y2 = 75)
Parameters:
<x1>,<y1>,
<x2>,<y2>
Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
Manual operation: See "Single Zoom" on page 448
DISPlay[:WINDow<n>]:ZOOM:STATe <State>
This command turns the zoom on and off.