Measurements
R&S
®
FSW
239User Manual 1173.9411.02 ─ 19
ally measured across a wide frequency range. The Spurious Emissions measurement
allows a flexible definition of all parameters. A result table indicates the largest devia-
tions of the absolute power from the limit line for each range, and the results can be
checked against defined limits automatically.
5.7.2 Spurious Emissions Measurement Results
The measured signal, including any spurious emissions, and optionally the detected
peaks are displayed in the Spurious Emissions measurement diagram. If defined, the
limit lines and the limit check results are also indicated. In addition to the graphical
results, a result table can be displayed to evaluate the measured powers and limit
check results (see also chapter 5.7.3.2, "Limit Lines in Spurious Measurements",
on page 241). The details of the evaluation list can be configured.
The following information is provided in the evaluation list for each range:
Column Description
Range Low Frequency range start for the range the peak value belongs to
Range Up Frequency range end for the range the peak value belongs to
RBW RBW of the range
Frequency Frequency at the peak value
Power Abs Absolute power level at the peak value
ΔLimit
Deviation of the absolute power level from the defined limit for the peak
value
Spurious Emissions Measurement

Table of Contents

Other manuals for Rohde & Schwarz FSW26