Common Analysis and Display Functions
R&S
®
FSW
477User Manual 1173.9411.02 ─ 19
Set to Default
Sets the color mapping to the default settings.
Remote command:
DISPlay[:WINDow<n>]:SGRam|SPECtrogram:COLor:DEFault on page 917
7.3.3 How to Configure Traces
The following step-by-step procedures describe the following tasks:
How to Configure a Standard Trace......................................................................477
How to Display and Configure a Spectrogram......................................................478
How to Copy Traces..............................................................................................482
7.3.3.1 How to Configure a Standard Trace
Step-by-step instructions on configuring the trace settings are provided here. For
details on individual functions and settings see chapter 7.3.2.1, "Trace Settings",
on page 464.
The remote commands required to perform these tasks are described in chap-
ter 11.8.2, "Configuring the Trace Display and Retrieving Trace Data", on page 909.
Trace settings are configured in the "Traces" dialog box.
To display the "Traces" dialog box, do one of the following:
Press the TRACE key and then select the "Trace Config" softkey.
Select "Analysis" from the "Overview", then select the "Traces" tab.
1. For each trace, select the "Trace Mode" and "Trace Detector". Traces with the
trace mode "Blank" are not displayed.
2. To configure several traces to predefined display modes in one step, press the but-
ton for the required function:
"Preset All Traces"
"Set Trace Mode Avg | Max | Min"
"Set Trace Mode Max | ClrWrite | Min"
For details see chapter 7.3.2.1, "Trace Settings", on page 464.
3. For "Average" trace mode, define the number of sweeps to be averaged in the
"Sweep/Average Count" field of the "Sweep Config" dialog box.
(Press the SWEEP key and then select the "Sweep Config" softkey to display the
"Sweep Config" dialog box.)
4. If linear scaling is used, select the "Average Mode: Linear".
5. To improve the trace stability, increase the number of "Sweep Points" or the
"Sweep Time".
All configured traces (not set to "Blank") are displayed after the next sweep.