Network and Remote Operation
R&S
®
FSW
655User Manual 1173.9411.02 ─ 19
"Off"
No wideband is used.
The option is indicated as "B7J".
"40 MHz"
The 40 MHz wideband is used.
The option is indicated as "B7J, 140".
"80 MHz"
The 80 MHz wideband is used.
The option is indicated as "B7J, 122".
Remote command:
SYSTem:PSA:WIDeband on page 1051
FSU/FSQ Preamplifier
This setting defines which option is returned when the *OPT? query is executed,
depending on the used preamplifier.
It is only available for FSU/FSQ emulation, and only if a preamplifier is used by the
R&S FSW (-B24/-B25 option).
Remote command:
SYSTem:PREamplifier on page 1051
Revision String
Defines the response to the REV? query for the revision number (HP emulation only,
see "GPIB Address" on page 651). Max. 36 characters are allowed.
Remote command:
SYSTem:REVision[:STRing] on page 1052
Resetting the Factory Revision
Resets the response to the REV? query for the revision number to the factory default
(HP emulation only, see "Language" on page 653).
Remote command:
SYSTem:REVision:FACTory on page 1052
10.4.4 LXI Settings
On the R&S FSW the LXI Class C functionality is already installed and enabled; thus,
the instrument can be accessed via any web browser (e.g. the Microsoft Internet
Explorer) to perform the following tasks:
●
modifying network configurations
●
modifying device configurations
●
monitoring connections from the device to other devices
The "LXI" tab of the "Network + Remote" dialog box provides basic LXI functions and
information for the R&S FSW.
Alternatively, you can change the LAN settings using the LXI Web browser interface.
For details see chapter 10.5.1.4, "How to Configure the LAN Using the LXI Browser
Interface", on page 661.
Only user accounts with administrator rights are able to use LXI functionality.
Network and Remote Control Settings