Remote Commands
R&S
®
FSW
1051User Manual 1173.9411.02 ─ 19
Note: For PSA89600 emulation, the option is indicated as "B7J" for the *OPT? query
("B7J, 140" if Wideband is activated).
Parameters:
<Language> "SCPI" | "8560E" | "8561E" | "8562E" | "8563E" | "8564E" |
"8565E" | "8566A" | "8566B" | "8568A" | "8568A_DC" | "8568B" |
"8568B_DC" | "8591E" | "8594E" | "71100C" | "71200C" |
"71209A" | "PSA89600" | "FSP" | "FSU" | "FSQ" | "FSV" |
"FSEA" | "FSEB" | "FSEM" | "FSEK"
*RST: SCPI
Example:
SYST:LANG '8560E'
Sets the system language to 8560E to simulate the HP model.
Usage: SCPI confirmed
Manual operation: See "Language" on page 653
SYSTem:PREamplifier <Option>
This setting defines which option is returned when the *OPT? query is executed,
depending on the used preamplifier.
It is only available for FSU/FSQ emulation, and only if a preamplifier is used by the
R&S FSW (-B24/-B25 option).
Parameters:
<Option> B23 | B24
*RST: B23
Manual operation: See "FSU/FSQ Preamplifier" on page 655
SYSTem:PSA:WIDeband <State>
This command defines which option is returned when the *OPT? query is executed,
depending on the state of the wideband option.
It is only available for PSA89600 emulation.
Parameters:
<State> ON | OFF | HIGH
OFF
The option is indicated as "B7J"
ON
The 40 MHz wideband is used.
The option is indicated as "B7J, 140".
HIGH
The 80 MHz wideband is used.
The option is indicated as "B7J, 122".
*RST: OFF
Emulating Other Instruments' Commands