Measurements
R&S
®
FSW
302User Manual 1173.9411.02 ─ 19
Remote command:
Marker selected via suffix <m> in remote commands.
5.13.4.2 EMI Final Measurement Configuration
The final EMI measurement can be performed with different settings than the initial
peak search. These settings are described here.
The detector to be used for the final EMI test can be defined differently for each fre-
quency, thus the detector is configured in the EMI marker settings, see "Final Test
Detector" on page 301.
Filter Type................................................................................................................... 302
RBW............................................................................................................................303
Automatic Peak Search...............................................................................................303
Dwell Time.................................................................................................................. 303
Frequency Axis Scaling...............................................................................................303
Res BW CISPR...........................................................................................................304
Res BW MIL................................................................................................................304
Filter Type
Defines the filter type.
The following filter types are available:
●
Normal (3dB)
●
Channel
●
RRC
●
5-Pole (not available for sweep type "FFT")
● CISPR (6 dB) - requires EMI (R&S FSW-K54) option
● MIL Std (6 dB) - requires EMI (R&S FSW-K54) option
For more information see chapter 6.5.1.6, "Which Data May Pass: Filter Types",
on page 406.
Electromagnetic Interference (EMI) Measurement (R&S FSW-K54)