Loading...
Measurements
R&S
®
FSW
273User Manual 1173.9411.02 ─ 19
5.10.3 Harmonic Distortion Results
As a result of the harmonics distortion measurement, the zero span sweeps of all
detected harmonics are shown in the diagram, separated by red display lines. This pro-
vides a very good overview of the measurement.
In addition, a result table is displayed providing the following information:
First harmonic frequency
THD (total harmonic distortion), relative and absolute values
For each detected harmonic:
Frequency
RBW
Power
Remote commands
The results can also be queried using remote commands.
The first harmonic frequency can be read out via the general center frequency com-
mand [SENSe:]FREQuency:CENTer on page 831.
THD: CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
on page 797
List of harmonics: CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST?
on page 798
5.10.4 Harmonic Distortion Configuration
Harmonic Distortion measurements are selected via the "Harmonic Distortion" button in
the "Select Measurement" dialog box. The measurement is started immediately with
the default settings. It can be configured via the MEAS CONFIG key or in the "Har-
Harmonic Distortion Measurement