5.6.7 Measurement Example: Multi-SEM Measurement
The following measurement example demonstrates an SEM measurement for a signal
with multiple sub blocks.
Test setup:
Signal
Generator 1
Coupler
[- 6 dB]
R&S FSW
Signal
Generator 2
Signal generator settings (e.g. R&S FSW SMU):
Device Standard Center Frequency Level Test Model
SigGen 1 3GPP/FDD 900 MHz 0 dBm 1-16
SigGen 2 EUTRA/LTE 906.5 MHz 0 dBm 1_1_5MHz
Setting up the measurement
1.
Preset the R&S FSW.
2. Set the center frequency to 903.25 MHz.
3. Set the reference level to 10 dBm with an offset of to 30 dB.
4. Press the MEAS key and select the "SEM" measurement function from the "Select
Measurement" dialog box.
Spectrum Emission Mask (SEM) Measurement