Remote Commands
R&S
®
FSW
836User Manual 1173.9411.02 ─ 19
11.7.2 Configuring Bandwidth and Sweep Settings
The commands required to configure the bandwidth, sweep and filter settings in a
remote environment are described here. The tasks for manual operation are described
in chapter 6.5, "Bandwidth, Filter and Sweep Configuration", on page 402.
Configuring the Bandwidth and Filter....................................................................836
Configuring the Sweep..........................................................................................839
11.7.2.1 Configuring the Bandwidth and Filter
[SENSe:]BANDwidth|BWIDth[:RESolution]....................................................................... 836
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO..............................................................836
[SENSe:]BANDwidth|BWIDth[:RESolution]:RATio..............................................................837
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE.............................................................. 837
[SENSe:]BANDwidth|BWIDth:VIDeo................................................................................ 838
[SENSe:]BANDwidth|BWIDth:VIDeo:AUTO.......................................................................838
[SENSe:]BANDwidth|BWIDth:VIDeo:RATio.......................................................................839
[SENSe:]BANDwidth|BWIDth:VIDeo:TYPE....................................................................... 839
[SENSe:]BANDwidth|BWIDth[:RESolution] <Bandwidth>
This command defines the resolution bandwidth and decouples the resolution band-
width from the span.
In the Realtime application, the resolution bandwidth is always coupled to the span.
For statistics measurements, this command defines the demodulation bandwidth.
For measurements on I/Q data in the frequency domain, the maximum RBW is 1 MHz.
Parameters:
<Bandwidth> refer to data sheet
*RST: RBW: AUTO is set to ON; DBW: 3MHz
Example:
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
Usage: SCPI confirmed
Manual operation: See "Analysis Bandwidth" on page 256
See "RBW" on page 303
See "Res BW CISPR" on page 304
See "Res BW MIL" on page 304
See "RBW" on page 427
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO <State>
This command couples and decouples the resolution bandwidth to the span.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Setting Basic Measurement Parameters

Table of Contents