Remote Commands
R&S
®
FSW
838User Manual 1173.9411.02 ─ 19
Parameters:
<FilterType> CFILter
Channel filters
NORMal
Gaussian filters
P5
5-pole filters
The 5-pole filter is not available for FFT sweeps.
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S FSW-K54) option
Return value for query is always PULS.
MIL
MIL Std (6 dB) - requires EMI (R&S FSW-K54) option
*RST:
NORMal
Example:
BAND:TYPE NORM
Manual operation: See "Filter Type" on page 302
See "Res BW CISPR" on page 304
See "Res BW MIL" on page 304
[SENSe:]BANDwidth|BWIDth:VIDeo <Bandwidth>
This command defines the video bandwidth.
The command decouples the video bandwidth from the resolution bandwidths.
Parameters:
<Bandwidth> refer to data sheet
*RST: AUTO is set to ON
Example:
BAND:VID 10 kHz
Usage: SCPI confirmed
Manual operation: See "VBW" on page 410
[SENSe:]BANDwidth|BWIDth:VIDeo:AUTO <State>
This command couples and decouples the video bandwidth to the resolution band-
width.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
BAND:VID:AUTO OFF
Usage: SCPI confirmed