General Instrument Setup
R&S
®
FSW
602User Manual 1173.9411.02 ─ 19
9.6.1 How to Perform a Self Test
The self test does not need to be repeated every time the instrument is switched on. It
is only necessary when instrument malfunction is suspected.
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
1. Press the SETUP key.
2. Press the "Service" softkey.
3. Press the "Selftest" softkey.
Once the instrument modules have been checked successfully, a message is dis-
played.
9.6.2 How to Align the Instrument
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
To perform a self-alignment
1. Press the SETUP key.
2. Select the "Alignment" softkey.
3. Select the "Start Self-alignment" button.
4. To abort the self-alignment process, select the "Abort Self-alignment" button.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
Press the SETUP key.
Press the "Alignment" softkey.