Measurements
R&S
®
FSW
244User Manual 1173.9411.02 ─ 19
For details on filter types see chapter 6.5.1.6, "Which Data May Pass: Filter Types",
on page 406.
Remote command:
[SENSe:]LIST:RANGe<range>:FILTer:TYPE on page 768
RBW
Sets the RBW value for this range.
For details on the RBW see chapter 6.5.1.1, "Separating Signals by Selecting an
Appropriate Resolution Bandwidth", on page 403.
Remote command:
[SENSe:]LIST:RANGe<range>:BANDwidth[:RESolution] on page 766
VBW
Sets the VBW value for this range.
For details on the VBW see chapter 6.5.1.2, "Smoothing the Trace Using the Video
Bandwidth", on page 404.
Remote command:
[SENSe:]LIST:RANGe<range>:BANDwidth:VIDeo on page 766
Sweep Time Mode
Activates or deactivates the auto mode for the sweep time.
For details on the sweep time mode see chapter 6.5.1.7, "How Long the Data is Mea-
sured: Sweep Time", on page 407
Remote command:
[SENSe:]LIST:RANGe<range>:SWEep:TIME:AUTO on page 773
Sweep Time
Sets the sweep time value for the range.
For details on the sweep time see chapter 6.5.1.7, "How Long the Data is Measured:
Sweep Time", on page 407
Remote command:
[SENSe:]LIST:RANGe<range>:SWEep:TIME on page 772
Detector
Sets the detector for the range.
For details refer to chapter 7.3.1.1, "Mapping Samples to Sweep Points with the Trace
Detector", on page 452.
Remote command:
[SENSe:]LIST:RANGe<range>:DETector on page 767
Ref. Level
Sets the reference level for the range.
Spurious Emissions Measurement