EasyManua.ls Logo

Rohde & Schwarz FSW26 - Measuring the Occupied Bandwidth

Rohde & Schwarz FSW26
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
727User Manual 1173.9411.02 ─ 19
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel<ch>]
[SENSe:]POWer:ACHannel:PRESet
Programming example: Measuring the carrier-to-noise ratio
This programming example demonstrates how to perform a Carrier-to-noise measure-
ment in a remote environment.
//----------Preparing the measurement------------
*RST
//Reset the instrument
FREQ:CENT 800MHz
//Sets the center frequency to the carrier frequency of 800 MHz.
CALC:MARK:FUNC:POW:SEL CN
//Activates carrier-to-noise ratio measurement.
POW:ACH:PRES CN
//Optimizes the instrument settings according to the channel bandwidth.
POW:ACH:PRES:RLEV
//Determines the ideal reference level for the measurement.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT;*WAI
//Initiates a new measurement and waits until the sweep has finished.
// Now turn off the carrier signal and repeat the measurement:
INIT;*WAI
//Initiates a new measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:POW:RES? CN
//Returns the carrier-to-noise ratio.
11.5.5 Measuring the Occupied Bandwidth
All remote control commands specific to occupied bandwidth measurements are
described here.
Configuring the Measurement...............................................................................727
Programming Example: OBW Measurement........................................................728
11.5.5.1 Configuring the Measurement
The following commands configure measurements of the occupied bandwidth.
Configuring and Performing Measurements