728User Manual 1173.9411.02 ─ 19
Useful commands for occupied bandwidth measurements described elsewhere
Configuring the channel:
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel<ch>]
[SENSe:]POWer:ACHannel:PRESet
[SENSe:]POWer:ACHannel:PRESet:RLEVel
Defining search limits:
CALCulate:MARKer:X:SLIMits[:STATe] on page 935
CALCulate:MARKer:X:SLIMits:LEFT on page 935
CALCulate:MARKer:X:SLIMits:RIGHT on page 935
Performing the measurement:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect on page 698
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe] on page 698
Retrieving results:
CALCulate:MARKer:FUNCtion:POWer<sb>:RESult? on page 696
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect on page 698
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe] on page 698
Remote commands exclusive to occupied bandwidth measurements
[SENSe:]POWer:BANDwidth|BWIDth...............................................................................728
[SENSe:]POWer:BANDwidth|BWIDth <Percentage>
This command selects the percentage of the total power that defines the occupied
bandwidth.
Parameters:
<Percentage> Range: 10 PCT to 99.9 PCT
*RST: 99 PCT
Example:
POW:BWID 95PCT
Manual operation: See "% Power Bandwidth" on page 191
11.5.5.2 Programming Example: OBW Measurement
This programming example demonstrates the measurement example described in
chapter 5.5.5, "Measurement Example", on page 193 in a remote environment.
//-------------Configuring the Measurement-----------------------
*RST
//Resets the instrument
FREQ:CENT 800MHz
//Sets the center frequency to 800 MHz.
DISP:TRAC:Y:RLEV -10dBm
//Sets the reference level to -10 dBm.
Configuring and Performing Measurements

Table of Contents

Other manuals for Rohde & Schwarz FSW26