Supported 89600 commands
:TRIGger[:SEQuence]:IF:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SOURce IMMediate|VIDeo|EXTernal<1|2>
:TRIGger[:SEQuence]:VIDeo:LEVel <numeric>
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency <freq>
11.14 Commands for Compatibility
The following commands are provided for compatibility to other signal analyzers only.
For new remote control programs use the specified alternative commands.
CALCulate:LIMit<k>:TRACe..........................................................................................1085
DISPlay[:WINDow<n>]:STATe.......................................................................................1085
DISPlay[:WINDow<n>]:TYPE........................................................................................ 1086
SOURce:EXTernal:FREQuency:SWEep[:STATe]............................................................1086
SYSTem:COMPatible...................................................................................................1086
TRIGger[:SEQuence]:BBPower:HOLDoff........................................................................1087
TRIGger[:SEQuence]:RFPower:HOLDoff........................................................................1087
CALCulate:LIMit<k>:TRACe <TraceNumber>
This command links a limit line to one or more traces.
Note that this command is maintained for compatibility reasons only. Limit lines no lon-
ger need to be assigned to a trace explicitely. The trace to be checked can be defined
directly (as a suffix) in the new command to activate the limit check (see CALCulate:
LIMit<k>:TRACe<t>:CHECk on page 986).
Parameters:
<TraceNumber> 1 to 6
*RST: 1
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
DISPlay[:WINDow<n>]:STATe <State>
This command changes the display state of the selected measurement window.
Note that this command is maintained for compatibility reasons only. Use the LAYout
commands for new remote control programs
(See chapter 11.6.2, "Working with Windows in the Display", on page 822).