Remote Commands
R&S
®
FSW
844User Manual 1173.9411.02 ─ 19
Example:
BAND:FFT NARR
Select narrow partial span for FFT filter.
11.7.3 Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
Amplitude Settings................................................................................................ 844
Configuring the Attenuation...................................................................................846
Configuring a Preamplifier.....................................................................................847
Scaling the Y-Axis.................................................................................................848
11.7.3.1 Amplitude Settings
The tasks for manual configuration are described in chapter 6.4.2, "Amplitude Set-
tings", on page 395.
Useful commands for amplitude configuration described elsewhere:
[SENSe:]ADJust:LEVel on page 863
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence...........................................................844
CALCulate<n>:UNIT:POWer...........................................................................................844
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel............................................................ 845
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel:OFFSet................................................845
[SENSe:]POWer:NCORrection........................................................................................ 845
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Usage: Event
Manual operation: See "Reference Level = Marker Level" on page 507
CALCulate<n>:UNIT:POWer <Unit>
This command selects the unit of the y-axis.
The unit applies to all measurement windows.