Remote Commands
R&S
®
FSW
848User Manual 1173.9411.02 ─ 19
INPut:GAIN:STATe <State>
This command turns the preamplifier on and off.
The command requires option R&S FSW-B24.
For R&S FSW 26 or higher models, the input signal is amplified by 30 dB if the pream-
plifier is activated.
For R&S FSW 8 or 13 models, the preamplification is defined by INPut:GAIN[:
VALue].
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INP:GAIN:STAT ON
Switches on 30 dB preamplification.
Usage: SCPI confirmed
Manual operation: See "Preamplifier (option B24)" on page 399
INPut:GAIN[:VALue] <Gain>
This command selects the preamplification level if the preamplifier is activated
(INP:GAIN:STAT ON, see INPut:GAIN:STATe on page 848).
The command requires option R&S FSW-B24.
Parameters:
<Gain> 15 dB | 30 dB
The availability of preamplification levels depends on the
R&S FSW model.
R&S FSW8/13: 15dB and 30 dB
R&S FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST: OFF
Example:
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Usage: SCPI confirmed
Manual operation: See "Preamplifier (option B24)" on page 399
11.7.3.4 Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]........................................................................849
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE....................................................849
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE............................................................. 849
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition........................................................849
DISPlay[:WINDow<n>]:TRACe:Y:SPACing.......................................................................850
Setting Basic Measurement Parameters