Remote Commands
R&S
®
FSW
849User Manual 1173.9411.02 ─ 19
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe] <Range>
This command defines the display range of the y-axis.
Note that the command works only for a logarithmic scaling. You can select the scaling
with DISPlay[:WINDow<n>]:TRACe:Y:SPACing.
Parameters:
<Range> Range: 1 dB to 200 dB
*RST: 100 dB
Example:
DISP:TRAC:Y 110dB
Usage: SCPI confirmed
Manual operation: See "Range" on page 400
DISPlay[:
WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again.
Usage:
SCPI confirmed
Manual operation: See "Auto Scale Once" on page 401
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE <Mode>
This command selects the type of scaling of the y-axis.
When the display update during remote control is off, this command has no immediate
effect.
Parameters:
<Mode> ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST: ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation: See "Scaling" on page 401
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition <Position>
This command defines the vertical position of the reference level on the display grid.
The R&S FSW adjusts the scaling of the y-axis accordingly.
For measurements with the external generator (R&S FSW-B10) the command defines
the position of the reference value.
Setting Basic Measurement Parameters