Common Measurement Settings
R&S
®
FSW
400User Manual 1173.9411.02 ─ 19
To configure the y-axis scaling settings
Vertical Axis settings can be configured via the AMPT key or in the "Amplitude" dialog
box.
To display the "Amplitude" dialog box, do one of the following:
Select "Amplitude" from the "Overview".
Select the AMPT key and then the "Scale Config" softkey.
The remote commands required to define these settings are described in chap-
ter 11.7.3, "Configuring the Vertical Axis (Amplitude, Scaling)", on page 844.
Range..........................................................................................................................400
Ref Level Position....................................................................................................... 400
Auto Scale Once......................................................................................................... 401
Scaling........................................................................................................................ 401
Range
Defines the displayed y-axis range in dB.
The default value is 100 dB.
Remote command:
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe] on page 849
Ref Level Position
Defines the reference level position, i.e. the position of the maximum AD converter
value on the level axis in %, where 0 % corresponds to the lower and 100 % to the
upper limit of the diagram.
Amplitude and Vertical Axis Configuration