EasyManua.ls Logo

Rohde & Schwarz FSW26 - Analyzing Statistics (APD, CCDF)

Rohde & Schwarz FSW26
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
778User Manual 1173.9411.02 ─ 19
CALC:PSE:SUBR 10
//Sets 10 peaks per range to be stored in the list.
//--------------Performing the Measurement-----
INIT:SPUR; *WAI
//Performs a spurious emission measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:LIM1:FAIL?
//Queries the result of the check for limit line 1.
TRAC? SPUR
//Queries the peak list of the spurious emission measurement.
11.5.8 Analyzing Statistics (APD, CCDF)
All remote control commands specific to statistical measurements are described here.
Activating Statistical Measurements..................................................................... 778
Configuring Statistical Measurements...................................................................779
Using Gate Ranges for Statistical Measurements................................................ 780
Scaling the Diagram..............................................................................................782
Performing a Statistical Measurement.................................................................. 784
Retrieving Results.................................................................................................784
Programming Example: Measuring Statistics....................................................... 786
11.5.8.1 Activating Statistical Measurements
The following commands activate statistical measurements.
CALCulate<n>:STATistics:APD[:STATe].......................................................................... 778
CALCulate<n>:STATistics:CCDF[:STATe]........................................................................778
CALCulate<n>:STATistics:APD[:STATe] <State>
This command turns the APD measurement on and off.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:STAT:APD ON
Switches on the APD measurement.
Manual operation: See "APD" on page 113
CALCulate<n>:STATistics:CCDF[:STATe] <State>
This command turns the CCDF on and off.
Configuring and Performing Measurements