The secure user mode setting is available in the general system configuration settings
(see chapter 9.3.4.5, "General Configuration Settings", on page 581).
Remote control
Initially after installation of the R&S FSW-K33 option, secure user mode must be
enabled manually once before remote control is possible. This is necessary to prompt
for a change of passwords.
8.3 Storing and Recalling Instrument Settings and Mea-
surement Data
Possibly you would like to restore or repeat a measurement you performed under spe-
cific conditions on the instrument. Or you want to evaluate imported data in another
application on the R&S FSW and would like to restore the measurement settings
applied during measurement. In these cases, you can store and recall instrument and
measurement settings, and possibly other related measurement data.
Two different methods are available for managing instrument settings:
Quick Save/Quick Recall - a defined set of instrument or channel settings are
stored or recalled quickly in just one step
Configurable Save/Recall - a user-defined set of instrument or channel settings are
stored to a definable storage location
Restrictions when recalling measurement settings
When recalling a saved configuration file, the following restrictions apply:
The frequency range defined in the configuration file must be supported by the
R&S FSW on which the file is to be recalled.
Configuration files created on a R&S FSW with certain options in use will not work
on an R&S FSW without these options
Files created with newer firmware versions may not work with a previous version
Saving instrument settings in secure user mode
Be sure to store instrument settings that you require beyond the current session before
Secure User Mode is enabled; see chapter 3.1.7, "Protecting Data Using the Secure
User Mode", on page 37.
Settings that are saved via QuickSave in secure user mode are only available during
the current session. As soon as the power is switched off on the R&S FSW, the data is
cleared.
Quick Save/Quick Recall.......................................................................................543
Configurable Storage and Recall.......................................................................... 545
How to Save and Load Instrument Settings..........................................................551
Storing and Recalling Instrument Settings and Measurement Data

Table of Contents

Other manuals for Rohde & Schwarz FSW26

Related product manuals